Instrument questions look easy, and for that reason they are a steady source of lost marks. A student sees a vernier scale, reads off two numbers, adds them, and writes an answer that is wrong by exactly the zero error, or wrong by the sign of the zero error, or wrong because the least count was computed from the wrong pair of scale divisions. The physics is one idea, that a vernier or a screw gauge reads a length to a fraction of the smallest division of its main scale, but the questions turn on a handful of conventions that must be applied in the right order.
The chapter guide, Dimensional Analysis and Error Analysis for JEE: Shortcuts, Checks and Traps, has one worked vernier reading among its ten examples and treats errors in general. This post takes the instruments themselves further: how the least count of each is found, every kind of zero error and its correction, how to read a screw gauge whose pitch is not $1$ mm, how to propagate the least count into a result, and how the same ideas apply to a stopwatch and a metre rule. The overview of the practical side of the syllabus is in Experimental Skills in JEE Main Physics. Every example below was checked with a script before it was written.
What this post covers
Measurement, instruments and errors are in the JEE Main syllabus under Units and Measurements and in the experimental section.
| Topic | What you must be able to do |
|---|---|
| Least count | Compute it for a vernier, a screw gauge and a stopwatch from the scale data |
| Reading | Combine main-scale and secondary-scale readings correctly |
| Zero error | Recognise positive and negative, and apply the correction with the right sign |
| Screw gauge | Find pitch and least count, read the linear and circular scales |
| Repeated readings | Mean, absolute error, mean absolute error, relative error |
| Error propagation | Carry the least count into areas, volumes, resistivity and $g$ |
| Reporting | Give the result with the correct number of significant figures |
The idea behind a vernier scale
A main scale is marked in equal divisions, each of length $s$ (for instance $1$ mm). A vernier scale slides along it and has $N$ divisions that together span $N-1$ main-scale divisions: $$N\cdot v = (N-1)\,s \quad\Longrightarrow\quad v = s - \frac{s}{N}.$$ Each vernier division is therefore shorter than a main-scale division by $s/N$. The least count (the smallest length the instrument can read) is the difference between one main-scale division and one vernier division: $$\text{LC} = s - v = \frac{s}{N} = \frac{\text{value of one main-scale division}}{\text{number of vernier divisions}}.$$ A scale with $s = 1$ mm and $N = 10$ has LC $= 0.1$ mm $= 0.01$ cm. With $N = 20$ it is $0.05$ mm, and a scale with $s = 0.5$ mm and $N = 25$ gives $0.02$ mm (Example 2 shows this).
The second type of vernier has the vernier divisions slightly longer than the main-scale divisions, for example $10$ vernier divisions equal to $11$ main-scale divisions. The least count is then the magnitude of the difference, $|v - s|$, which here equals $0.1s$. Every question on this should be settled by taking the difference of the two division lengths, and not by remembering a formula.
Reading a vernier
The reading has two parts. $$\text{Reading} = \text{MSR} + n\times\text{LC},$$ where MSR is the main-scale reading, the last main-scale mark to the left of the vernier zero, and $n$ is the number of the vernier division that coincides with a main-scale division. Write both in the same unit before you add. The measured length is the length between the jaws, taken after correcting for zero error.
Also read: Experiments in Measurement NEET PYQs: Vernier, Screw Gauge
Zero error on a vernier
With the jaws closed the vernier zero should coincide with the main-scale zero. If it does not, there is a zero error, and every reading carries it.
- Positive zero error: the vernier zero is to the right of the main zero. If the $n$-th vernier division coincides, the zero error is $+n\times\text{LC}$. The corrected reading is reading minus zero error, so it is smaller than the observed value.
- Negative zero error: the vernier zero is to the left of the main zero. Here the coinciding division is counted from the right end, so with $N$ divisions on the vernier and the $n$-th division coinciding, the zero error is $-(N - n)\times\text{LC}$. The corrected reading is the observed value plus the magnitude of the error, so it is larger.
The rule that never changes is $$\text{corrected reading} = \text{observed reading} - \text{zero error},$$ with the zero error taken with its sign. For a positive error you subtract, for a negative error you add. The negative case is where most mistakes occur: students read the coinciding division as the error without subtracting it from $N$ (Example 3).
The screw gauge
A screw gauge measures small lengths such as the diameter of a wire. It has a main (linear) scale on the barrel and a circular scale with $N$ divisions on the thimble. When the thimble makes one complete rotation it advances along the barrel by the pitch $p$: $$\text{pitch} = \frac{\text{distance moved by the screw}}{\text{number of complete rotations}}.$$ The least count is the distance advanced per circular-scale division: $$\text{LC} = \frac{\text{pitch}}{\text{number of circular divisions}} = \frac{p}{N}.$$ Typical numbers are $p = 0.5$ mm with $N = 50$ (LC $= 0.01$ mm) and $p = 1$ mm with $N = 100$ (LC $= 0.01$ mm). Do not assume that the pitch equals $1$ mm: questions often give the distance moved in several rotations, and the pitch must be found from that (Example 4).
Reading a screw gauge
$$\text{Reading} = \text{linear scale reading} + (\text{circular scale division at the reference line})\times\text{LC}.$$ The linear scale reading is the last visible mark on the barrel to the left of the thimble edge. If the pitch is $0.5$ mm the barrel has marks every $0.5$ mm, with those on one side of the reference line marking half-millimetres. Be careful to count a half-millimetre mark when it is visible.
Zero error on a screw gauge
Close the gap gently by turning the ratchet and read the circular scale at the reference line.
- If the circular scale reads a small number $n$ at the reference line (for example $3$, with the linear scale at zero), the zero error is $+n\times\text{LC}$, positive.
- If the circular scale reads a large number, such as $47$ out of $50$, the zero has gone past the reference line and the error is negative: it is $-(N - n)\times\text{LC} = -3\times\text{LC}$.
As with the vernier, the corrected reading is the observed reading minus the zero error with its sign (Example 5 does one of each).
From least count to error
The least count is the smallest quantity the instrument can resolve, and the usual convention in JEE is that the absolute error of a single reading from that instrument equals the least count. For repeated readings the absolute error of the mean is usually the mean of the absolute deviations. The relative error is the absolute error divided by the measured value, and a percentage error is that times $100$.
Then the standard propagation rules apply:
- For a sum or difference, absolute errors add.
- For a product or quotient, relative errors add.
- For a power $x^n$, the relative error is multiplied by $|n|$.
For the area of a circle from a measured diameter $d$, $A = \pi d^2/4$, the relative error is $2\,\Delta d/d$, twice that of the diameter. For the volume of a sphere, $V = \pi d^3/6$, the relative error is $3\,\Delta d/d$. A small diameter measured with a coarse instrument has a large relative error, and a squared or cubed dependence makes it larger. This is the reason a thin wire is measured with a screw gauge and not with a vernier.
Stopwatch and metre rule
The same least-count logic applies to the other instruments. A stopwatch reading to $0.1$ s has a least count of $0.1$ s. If you time $n$ oscillations of a pendulum, the error in the period is the least count divided by $n$, so timing $20$ oscillations makes the error in $T$ twenty times smaller than timing one. A metre rule marked in millimetres has LC $= 1$ mm, though reading an edge at both ends often means two readings and so a combined error; a question will say which to use.
Worked examples
Example 1 (numerical, easy)
A vernier callipers has the main scale in millimetres, and $20$ divisions of its vernier scale coincide with $19$ divisions of the main scale. Find the least count in mm. A reading is taken with a main-scale reading of $24$ mm and the $13$th vernier division coinciding with a main-scale division. Find the length in mm to two decimal places.
Solution. The main-scale division is $s = 1$ mm, and $N = 20$: $$\text{LC} = \frac{s}{N} = \frac{1}{20} = 0.05\ \text{mm}.$$ The length is $$24 + 13\times0.05 = 24 + 0.65 = 24.65\ \text{mm}.$$
Answers: $0.05$ mm; $24.65$ mm. A check on the arithmetic is that the vernier reading $0.65$ mm must be less than one main-scale division, $1$ mm. If it is not, the number of vernier divisions has been misread.
Example 2 (single-correct, easy)
On a vernier callipers each main-scale division is $0.5$ mm and $25$ vernier divisions coincide with $24$ main-scale divisions. The least count is
(A) $0.01$ mm (B) $0.02$ mm (C) $0.05$ mm (D) $0.5$ mm
Solution. $$\text{LC} = \frac{s}{N} = \frac{0.5}{25} = 0.02\ \text{mm}.$$
Check by computing the vernier division directly: $v = 24\times0.5/25 = 0.48$ mm, and $s - v = 0.5 - 0.48 = 0.02$ mm.
Answer: (B). Option (D) is the size of the main-scale division. Option (C) comes from using $N = 10$ by habit, and option (A) from dividing $0.5$ mm by $50$. The definition, a difference of two division lengths, never fails.
Example 3 (numerical, medium)
A vernier callipers has $10$ vernier divisions equal to $9$ main-scale divisions, with the main scale marked in $1$ mm, so the least count is $0.01$ cm. When the jaws are closed, the zero of the vernier lies to the left of the main-scale zero and the $6$th vernier division coincides with a main-scale division. An object is measured and the observed reading is $3.42$ cm. Find the zero error in cm and the corrected length in cm. Then, for a different instrument with a positive zero error where the $4$th division coincides, find the corrected length for an observed reading of $2.86$ cm.
Solution. The first instrument has a negative zero error. The coinciding division counts from the right, so $$\text{zero error} = -(N - n)\times\text{LC} = -(10 - 6)\times0.01 = -0.04\ \text{cm}.$$ Corrected reading $=$ observed $-$ zero error: $$3.42 - (-0.04) = 3.46\ \text{cm}.$$
For the second instrument the error is $+4\times0.01 = +0.04$ cm, so $$2.86 - 0.04 = 2.82\ \text{cm}.$$
Answers: $-0.04$ cm and $3.46$ cm; $2.82$ cm. The tempting wrong answer in the first part is a zero error of $-0.06$ cm, from reading $n = 6$ as the error. That would give a corrected length of $3.48$ cm. The test is a sense of direction: a negative error means the instrument under-reads, so the true length must be bigger than the reading.
Example 4 (numerical, medium)
A screw gauge advances $2.5$ mm when its thimble is turned through $5$ complete rotations. The circular scale has $100$ divisions. Find the pitch in mm and the least count in mm. A wire is measured with the screw gauge. Find its diameter if the linear scale reading is $4$ mm and the circular scale reading is $36$, with a negative zero error of $-0.07$ mm. Give the diameter to two decimal places in mm.
Solution. $$\text{pitch} = \frac{2.5}{5} = 0.5\ \text{mm},\qquad \text{LC} = \frac{0.5}{100} = 0.005\ \text{mm}.$$
The observed reading is $4 + 36\times0.005 = 4.18$ mm. With the zero error $-0.07$ mm, $$\text{corrected} = 4.18 - (-0.07) = 4.25\ \text{mm}.$$
Answers: pitch $0.5$ mm; LC $0.005$ mm; diameter $4.25$ mm. The usual slip is to take the least count as $1/100$ mm, assuming a pitch of $1$ mm. That would give a reading of $4.36$ mm and a corrected diameter of $4.43$ mm, which is off by $0.18$ mm. Always compute the pitch from the data.
Example 5 (numerical, medium)
A screw gauge has a pitch of $0.5$ mm and $50$ divisions on the circular scale, so the least count is $0.01$ mm. With the jaws closed, the circular scale reads $+3$ divisions at the reference line (positive zero error). A sphere is measured and the linear scale shows $3$ mm and the circular scale reading is $27$. Find the corrected diameter in mm. Then repeat the calculation for an instrument whose closed-jaw circular reading is $47$ (negative zero error), with the same observed reading.
Solution. Observed reading: $3 + 27\times0.01 = 3.27$ mm.
Positive error: $+3\times0.01 = +0.03$ mm. Corrected: $3.27 - 0.03 = 3.24$ mm.
Negative error: the closed reading $47$ means the zero lies $50 - 47 = 3$ divisions past the reference line, so the error is $-3\times0.01 = -0.03$ mm. Corrected: $3.27 - (-0.03) = 3.30$ mm.
Answers: $3.24$ mm and $3.30$ mm. Both instruments gave the same observed reading, but opposite corrections. In the exam, write the zero error with its sign before you do anything else, and then apply "observed minus zero error".
Example 6 (numerical, medium)
The diameter of a wire is measured with a screw gauge of least count $0.01$ mm as $d = 1.50$ mm. Find the percentage error in the cross-sectional area of the wire, to two decimal places. Take the absolute error in $d$ to be the least count.
Solution. The area is $A = \pi d^2/4$, so $$\frac{\Delta A}{A} = 2\,\frac{\Delta d}{d} = 2\times\frac{0.01}{1.50} = 0.01333.$$ As a percentage this is $1.33\%$.
Answer: $1.33\%$. The error in the diameter alone is $0.67\%$; squaring doubles the relative error. If the same wire were measured with a vernier of least count $0.1$ mm, the error in $d$ would be $6.7\%$ and in the area $13.3\%$, which is the reason a screw gauge is chosen.
Example 7 (numerical, hard)
The resistivity of a wire is found from $\rho = \dfrac{\pi R d^2}{4L}$. The measurements are the resistance $R = 10.0 \pm 0.1\ \Omega$, the diameter $d = 0.50 \pm 0.01$ mm and the length $L = 50.0 \pm 0.1$ cm. Find the maximum percentage error in $\rho$.
Solution. The relative errors of the factors are
- $R$: $0.1/10.0 = 1.0\%$,
- $d$ (appearing squared): $2\times0.01/0.50 = 4.0\%$,
- $L$ (in the denominator, so the error still adds): $0.1/50.0 = 0.2\%$.
For a product and quotient the maximum relative error is the sum: $$\frac{\Delta\rho}{\rho} = 1.0\% + 4.0\% + 0.2\% = 5.2\%.$$
Answer: $5.2\%$. The diameter dominates because it is small and squared. In an experiment, the way to cut the total error is to improve the measurement of the diameter, for example by averaging several readings along the wire, and not the length.
Example 8 (numerical, medium)
Five readings of the diameter of a sphere with a vernier callipers (least count $0.01$ cm) are $2.34$, $2.36$, $2.35$, $2.33$ and $2.37$ cm. Find the mean, the mean absolute error, and the percentage error in the volume of the sphere, to two decimal places.
Solution. The mean is $$\bar d = \frac{2.34 + 2.36 + 2.35 + 2.33 + 2.37}{5} = 2.35\ \text{cm}.$$ The absolute deviations are $0.01$, $0.01$, $0.00$, $0.02$, $0.02$, so $$\overline{\Delta d} = \frac{0.01 + 0.01 + 0.00 + 0.02 + 0.02}{5} = 0.012\ \text{cm}.$$ The relative error in the diameter is $0.012/2.35 = 0.51\%$. For a sphere, $V\propto d^3$: $$\frac{\Delta V}{V} = 3\times0.51\% = 1.53\%.$$
Answers: mean $2.35$ cm; mean absolute error $0.012$ cm; percentage error in the volume $1.53\%$. The result is reported as $d = 2.35 \pm 0.01$ cm, with the error rounded to the same decimal place as the instrument allows. Quoting the mean to three decimal places would claim a precision that the instrument does not have.
Example 9 (single-correct, medium)
A vernier scale has $10$ divisions equal to $11$ main-scale divisions, and the main scale is marked in millimetres. The least count of this vernier is
(A) $0.1$ mm (B) $1.1$ mm (C) $0.11$ mm (D) $1$ mm
Solution. Each vernier division is $v = 11\times1/10 = 1.1$ mm. The least count is the magnitude of the difference between a vernier division and a main-scale division: $$\text{LC} = |v - s| = |1.1 - 1| = 0.1\ \text{mm}.$$
Answer: (A). The vernier divisions here are longer than the main-scale divisions, so the vernier reading counts in the opposite direction, but the least count has the same value, $s/N = 1/10 = 0.1$ mm. The formula $s/N$ does apply, with $N$ the number of vernier divisions in both cases.
Example 10 (numerical, hard)
A student times $20$ oscillations of a simple pendulum with a stopwatch of least count $0.1$ s and records $36.0$ s. The length, measured with a metre rule of least count $0.1$ cm, is $80.0$ cm. Find the period and its absolute error, then the percentage error in $g = 4\pi^2L/T^2$, to two decimal places. Take the absolute error of each quantity as its least count.
Solution. The period is $T = 36.0/20 = 1.80$ s. The error in the period is the stopwatch least count divided by $20$: $$\Delta T = \frac{0.1}{20} = 0.005\ \text{s}.$$
The percentage errors are $\Delta L/L = 0.1/80.0 = 0.125\%$ and $\Delta T/T = 0.005/1.80 = 0.278\%$. Since $g\propto LT^{-2}$, $$\frac{\Delta g}{g} = \frac{\Delta L}{L} + 2\frac{\Delta T}{T} = 0.125\% + 0.556\% = 0.68\%.$$ The value of $g$ is $4\pi^2\times0.800/(1.80)^2 = 9.75\ \text{m s}^{-2}$, so the absolute error is about $0.066\ \text{m s}^{-2}$.
Answers: $T = 1.80 \pm 0.005$ s; $\Delta g/g = 0.68\%$. Had the student timed a single oscillation, the error in $T$ would have been $0.1$ s, which is $5.6\%$ of a $1.8$ s period, and the error in $g$ would have been about eleven per cent. Timing many oscillations is the standard remedy.
Example 11 (multiple-correct, medium)
Which of the following statements about the zero error of a screw gauge and a vernier callipers are correct?
(A) A positive zero error means the instrument over-reads, so the zero error must be subtracted from the observed reading. (B) A negative zero error is added, in magnitude, to the observed reading. (C) For a vernier with $N$ divisions, a negative zero error with the $n$-th division coinciding has magnitude $(N - n)\times\text{LC}$. (D) The zero error changes the least count of the instrument.
Solution. (A) is correct. A positive error means the jaws read a positive length when closed, so every measurement is too large by that amount. (B) is correct, since "observed $-$ (negative error)" is observed plus the magnitude. (C) is correct, and is the rule used in Example 3. (D) is wrong. The zero error shifts every reading by the same amount but does not change the least count, which is fixed by the scale design.
Answer: (A), (B) and (C).
Common traps
- Computing the least count from the wrong pair of lengths. It is the difference between one main-scale division and one vernier division, which equals $s/N$ when $N$ vernier divisions span $N - 1$ main divisions. If a question gives a different relation, compute the vernier division from it and subtract.
- Reading the negative zero error as the coinciding division number. For negative error the magnitude is $(N - n)\times\text{LC}$. In Example 3 the coinciding division was the $6$th, but the magnitude is $4$ divisions.
- Adding a positive zero error. The correction is always observed minus zero error with the sign. Positive error: subtract. Negative error: add.
- Assuming the pitch is $1$ mm. The pitch is the distance moved per rotation. A screw gauge that advances $2.5$ mm in $5$ turns has a pitch of $0.5$ mm, and the least count follows from that.
- Mixing units when adding the two parts of a reading. The main-scale reading may be in cm and the vernier part in mm. Convert to a single unit before adding.
- Forgetting a visible half-millimetre mark on the linear scale. On a screw gauge of pitch $0.5$ mm the barrel carries marks on both sides of the reference line. A mark that is visible counts and adds $0.5$ mm to the linear reading.
- Applying the error of a single reading to a mean. If several readings are taken, the error to quote is the mean absolute deviation, not the least count alone, unless the question says to use the least count.
- Omitting the power in the error formula. The percentage error in the area of a circle is twice that of the diameter, and in a sphere's volume three times. Forgetting the index understates the error.
- Subtracting errors in a difference. The absolute errors of a sum and of a difference both add. A difference of two close readings has a large relative error.
- Quoting too many figures. A reading from an instrument of least count $0.01$ mm cannot be reported as $3.2437$ mm. Report to the decimal place of the least count and round the final result to the same number of significant figures as the least precise measurement.
- Timing a single oscillation. The relative error in $T$ falls as $1/n$ when $n$ oscillations are timed. A single oscillation with a stopwatch of $0.1$ s has an enormous relative error.
- Treating zero error as part of the least count. A zero error shifts all readings equally; it neither adds to nor reduces the instrument's resolution. Correct for it, then apply the least count to the error budget.
A one-page recap
| Idea | Result |
|---|---|
| Vernier least count | $\text{LC} = s/N$ (value of one main division over number of vernier divisions) |
| Vernier reading | $\text{MSR} + n\times\text{LC}$ |
| Positive zero error | $+n\times\text{LC}$; subtract it |
| Negative zero error | $-(N - n)\times\text{LC}$; add its magnitude |
| Correction rule | corrected = observed $-$ zero error (with sign) |
| Screw gauge pitch | distance moved per rotation = distance / number of rotations |
| Screw gauge LC | pitch / number of circular divisions |
| Screw gauge reading | linear reading $+$ circular reading $\times$ LC |
| Error of one reading | the least count |
| Mean absolute error | mean of $|x_i - \bar x|$ |
| Relative and percentage error | $\Delta x/x$, and times $100$ |
| Sum or difference | absolute errors add |
| Product or quotient | relative errors add |
| Power $x^n$ | relative error $\times|n|$ |
| Area of a circle | $2\,\Delta d/d$; sphere volume $3\,\Delta d/d$ |
| $n$ oscillations | $\Delta T = \text{LC}/n$ |
Also read: Young's Double Slit Problems in Telugu: JEE ఫ్రింజ్ లెక్కలు
How to practise this chapter
Read the Units and Measurements chapter of the book first, in the language you study best: English, Hindi or Telugu. Then use the chapter tests in sequence: Units and Measurements, test 1, test 2, test 3, and on to the seventh. The chapter has seven tests of fifteen questions each, and the instrument questions are mixed in with units, dimensions and significant figures. The free JEE Main mini mock is a good way to get used to the format. Subject practice tests and full mocks, part of the Pro Pass, come once the chapter tests are going well.
An eight-day plan, assuming you have already read the chapter guide:
| Day | Work |
|---|---|
| 1 | The vernier principle and least count; Examples 1, 2 and 9; draw a vernier with $N = 10$ and mark a reading by hand |
| 2 | Zero error on the vernier, positive and negative; Example 3; test 1 |
| 3 | The screw gauge: pitch, least count, reading; Example 4; test 2 |
| 4 | Screw gauge zero error; Examples 5 and 11; test 3 |
| 5 | Repeated readings and mean absolute error; Example 8; test 4 |
| 6 | Error propagation with instruments; Examples 6 and 7; test 5 |
| 7 | The pendulum and stopwatch error; Example 10; tests 6 and 7 |
| 8 | Redo every miss and write the recap table from memory; take a short mixed set |
Record each miss with its cause in the format of the error log and mock analysis method. For this topic the cause is nearly always one of three things: the sign of the zero error, the least count computed from the wrong data, or the power left out of an error formula. The wider study plan for the subject is in JEE Main Physics preparation strategy, and the numerical-value habits that these questions reward are in JEE Main numerical value questions. The nearby topic of units and dimensions is covered in Units, Measurement and Kinematics for JEE. Logged-in students can open the tests from their JEE Main dashboard.
Frequently asked questions
Why does the least count of a vernier equal $s/N$?
Because $N$ vernier divisions are made equal to $N - 1$ main-scale divisions, each vernier division is shorter than a main division by $s/N$. When the $n$-th vernier line coincides with a main-scale line, the vernier zero is ahead of the previous main-scale mark by $n\,s/N$, which is how the instrument reads fractions of a division.
How can I tell a positive from a negative zero error at a glance?
Close the jaws and look at the vernier zero relative to the main-scale zero. If it is to the right of it, the error is positive. If it is to the left, the error is negative. For a screw gauge, a small circular reading at the reference line is a positive error and a large reading, close to the full count, is a negative error.
Should I add or subtract the zero error?
Always subtract the zero error with its sign: corrected reading equals observed reading minus zero error. A positive error is subtracted, and a negative error is subtracted as a negative number, which adds its magnitude. Writing the rule as one line avoids having to remember two cases.
Is the pitch of a screw gauge always $1$ mm?
No. Pitches of $0.5$ mm and $1$ mm are both common, and questions often give the distance moved in a number of rotations so that you work out the pitch yourself. The least count is the pitch divided by the number of circular divisions.
What is the absolute error of a single reading?
By the usual JEE convention it is the least count of the instrument. For a set of readings the mean of the absolute deviations from the mean is used. A question will usually make clear which of the two is intended.
Why is a screw gauge used for thin wires and not a vernier callipers?
Its least count is smaller, typically $0.01$ mm against $0.1$ mm or $0.05$ mm for a vernier. For a thin wire the relative error is the least count divided by the diameter, so the finer instrument keeps the error small, and it is amplified when the diameter is squared to give the area.
Do I need the units of the circular and linear scales to be the same?
Yes when you add the two parts of the reading. Convert one to the unit of the other, normally to millimetres, before adding. The least count of a screw gauge is in the unit of its pitch.